It is an innovated portable battery densitometer designed for measuring the optical density (blackening) of technical radiographs.
MD20 digital densitometer
Advantages of the device
- very high accuracy thanks to the use of an integrated circuit with the ability to logarithm in four decades with an error of 0.2 %
- zeroing at any illumination
- one of the smallest densitometers on the market
- it is possible to measure a value of up to 4D with sufficient illumination
- favourable price
Technical parameters
measuring range 0 - 4D resolution 0,01D accuracy ±0,05D (0 - 4D) screen display optical density, +/-, BATT zero setting range in any light zero hold time approx 5 min, (± 0,01D) zero stabilization time max. 1s reading stabilization time max. 2s probe aperture diameter 2mm power supply – battery 9V typ IEC 6F22 or 6LR61 9V consumption measuring/off 11,0mA operating time (measurement) approx. 45 hrs. (alkaline battery) control 4 buttons: on/off, zero setting, switch to calibration mode, change value during calibration automatic shutdown approx. 45 s when idle after switching on ambient temperature 5 - 35 °C dimensions 79 x 118 x 25mm cable length 70cm weight 155g Standard delivery
device, probe on 70cm long cable, 9V battery, manual, transport case